Balance Electronic Analytical If your balance offers an internal calibration feature that requires minimal user input, follow the manufacturer’s instructions and use this feature regularly. However, if quantitative chemical analysis plays a critical role in your industry (e.g. food or pharma), conduct your own tests as well.

t’s a good idea to set up a regular calibration schedule, so your balances can be tested for accuracy every few months. Also, make sure to recalibrate them in case of any environmental changes (e.g. relocating the device).

The appropriate calibration settings for a certain environment can be determined with certified calibration masses, which is especially crucial when you’re calibrating sensitive instruments like analytical balances.

Physical Influences and Calibration

Analytical balance readouts can be affected by changes in the sample and/or container as well as other physical influences.

    Use clean and dry containers for samples, and avoid beakers with large or wide openings. Pick the smallest container possible, and cover it if needed.

    Condition the sample/container to the prevailing temperature, to avoid issues with repeatability and weighing results. You may also find drifts in the readout value (even with a stable display when the balance isn’t loaded) otherwise.

    Evaporation may occur in uncovered samples/containers, or they may absorb water if handled with bare fingers. Readouts could increase or decrease continuously, or you may find drift (even with a stable display when the balance isn’t loaded).

    Electrostatically charged samples/containers (e.g. glass or plastic) or very low humidity may cause non-repeatable values or drifts in one direction. Shield the sample with metal foil or a metal container, increase the humidity and neutralize static electricity with an ionizing blower.

    Magnetic force fields can lead to stable but non-repeatable values, or changes in display based on the sample’s position. Avoid using magnetic stirring bars, shield samples with Mumetal foil or non-magnetic objects, use anti-magnetic weighing pans and demagnetize the pan before weighing.

Thank you for allowing Accurate Weighing Scales (U) Ltd the privilege to serve you in advance.

For inquiries on deliveries contact us

Office +256 (0) 705 577 823, +256 (0) 775 259 917

Address: Wandegeya KCCA Market South Wing, 2nd Floor Room SSF 036

Email:  weighingscales@countrywinggroup.com

Balance Electronic Analytical
Balance Electronic Analytical